Patients undergoing a percutaneous bone biopsy often complain of pain during needle insertion, despite local anesthesia. Bone biopsy needles are typically inserted with combined axial and twisting motions. These motions could cause pain through frictional heating or direct mechanical irritation. The hypothesis of this study is that the insertion energy of bone biopsy needles can be reduced by modifying the insertion kinetics or by adding a friction-lowering coating to the needles. Jamshidi bone biopsy needles were driven into a bone analog model by an MTS materials testing machine operating under axial and rotational displacement control. The load/torque recordings showed that, to significantly decrease insertion energy and peak resistance to needle insertion, axial velocity and angular frequency had to be decreased to one quarter of the baseline, typical-usage parameters. However, the increased insertion time may not be acceptable clinically. The majority of the insertion energy was associated with the needle axial thrust rather than with needle twisting. Overcoming friction against the side of the needle consumed much more of the insertion energy than did the process of cutting per se. None of five needle coatings tested succeeded in appreciably lowering the insertion energy, and none achieved a substantial decrease in peak resisting force.
Skip Nav Destination
Article navigation
December 2001
Technical Papers
Frictional Insertion Kinetics of Bone Biopsy Needles
Anneliese D. Heiner,
Anneliese D. Heiner
Department of Orthopaedic Surgery, University of Iowa, Iowa City, IA 52242
Search for other works by this author on:
Thomas D. Brown,
Thomas D. Brown
Departments of Orthopaedic Surgery and Biomedical Engineering, University of Iowa, Iowa City, IA 52242
Search for other works by this author on:
Victor Rossin,
Victor Rossin
Allegiance Healthcare Corporation, McGaw Park, IL 60085
Search for other works by this author on:
Joseph A. Buckwalter
Joseph A. Buckwalter
Department of Orthopaedic Surgery, University of Iowa, Iowa City, IA 52242
Search for other works by this author on:
Anneliese D. Heiner
Department of Orthopaedic Surgery, University of Iowa, Iowa City, IA 52242
Thomas D. Brown
Departments of Orthopaedic Surgery and Biomedical Engineering, University of Iowa, Iowa City, IA 52242
Victor Rossin
Allegiance Healthcare Corporation, McGaw Park, IL 60085
Joseph A. Buckwalter
Department of Orthopaedic Surgery, University of Iowa, Iowa City, IA 52242
Contributed by the Bioengineering Division for publication in the JOURNAL OF BIOMECHANICAL ENGINEERING. Manuscript received by the Bioengineering Division January 11, 2001; revised manuscript received July 10, 2001. Associate Editor: C. H. Turner.
J Biomech Eng. Dec 2001, 123(6): 629-634 (6 pages)
Published Online: July 10, 2001
Article history
Received:
January 11, 2001
Revised:
July 10, 2001
Citation
Heiner, A. D., Brown, T. D., Rossin, V., and Buckwalter, J. A. (July 10, 2001). "Frictional Insertion Kinetics of Bone Biopsy Needles ." ASME. J Biomech Eng. December 2001; 123(6): 629–634. https://doi.org/10.1115/1.1407829
Download citation file:
Get Email Alerts
Cited By
Related Articles
Development of a Force-Driven Distractor for Distraction Osteogenesis
J. Med. Devices (December,2011)
On the Determination of the Angular Orientation of a Vertebra
J Biomech Eng (February,2002)
Response to Dr. Robert W. Mann
J Biomech Eng (August,2004)
Related Proceedings Papers
Related Chapters
Novel and Efficient Mathematical and Computational Methods for the Analysis and Architecting of Ultralight Cellular Materials and their Macrostructural Responses
Advances in Computers and Information in Engineering Research, Volume 2
Application of Non-Linear Elastic Wave Spectroscopy (NEWS) to In Vitro Damage Assessment in Cortical Bone
Biomedical Applications of Vibration and Acoustics in Imaging and Characterizations
Research on Oracle Bone Inscriptions Machine Translation Based on Example and Ontology
International Conference on Advanced Computer Theory and Engineering, 4th (ICACTE 2011)